2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-41
Details
This command selects which set of input and output terminals the instrument uses. You can select
front panel or rear panel terminals.
If the output is on when you change from one set of terminals to the other, the output is turned off.
Example
:ROUT:TERM REAR
:ROUT:TERM?
Set the instrument to use the rear-panel
terminals and query to verify.
Output:
REAR
Also see
None
SCRipt subsystem
The SCRipt subsystem controls macro or instrument setup scripts. For additional information on
macro scripts, refer to
Saving front-panel settings into a macro script
(on page 3-53).
:SCRipt:RUN
This command runs a script.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
SCRipt:RUN "<scriptName>"
<scriptName>
The name of the script
Details
The script must be available in the instrument to be used by this command.
Example
SCR:RUN "bufferCreate"
Runs a script named
bufferCreate
.
Also see
Saving front-panel settings into a macro script
(on page 3-53)
(on page 3-35)