2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-29
:DISPlay:<function>:DIGits
This command determines the number of digits that are displayed for measurements on the front panel.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Measure: 5
Usage
:DISPlay:<function>:DIGits <value>
:DISPlay:<function>:DIGits <DEF|MIN|MAX>
:DISPlay:<function>:DIGits?
:DISPlay:<function>:DIGits? <DEF|MIN|MAX>
<function>
The measure function:
Current:
CURRent[:DC]
Resistance:
RESistance
Voltage:
VOLTage[:DC]
<value>
Display digits:
6.5:
6
5.5:
5
4.5:
4
3.5:
3
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value
Details
This command affects how the reading for a measurement is displayed on the front panel of the
instrument. It does not affect the number of digits returned in a remote command reading. It also does
not affect the accuracy or speed of measurements.
The display digits setting is saved with the function setting, so if you use another function, then return
to the function for which you set display digits, the display digits setting you set previously is retained.
The change in digits occurs the next time a measurement is made.
To change the number of digits returned in a remote command reading, use
:FORMat:ASCii:PRECision
.
If you send this command without the
<function>
parameter, the digits values for all measure
functions are changed.
Example
:DISP:CURR:DIG 5
Set the front panel to display current measurements with 5½ digits.
Also see
(on page 12-34)