Model 8010 High Power Device Test Fixture User's Manual
Section 8: Frequently asked questions
8010-900-01 Rev. C / March 2017
8-3
Can I operate the test fixture with the lid open?
The lid of the Model 8010 test fixture isolates the operator from hazardous voltages that are
potentially present at the device under test and cable ports. The lid also functions as a blast shield,
protecting the user from flying debris and fire.
The interlock circuitry of the Model 8010 test fixture is designed to be engaged only when the lid is
closed. If the lid is open, the interlock is disengaged and any SourceMeter Instruments (SMUs) with
outputs that are configured for high voltage will not turn on.
Low voltage SMUs may still operate with the lid open and without the interlock connected. For
example, the Model 2651 will still operate if the output enable function is not activated. Additionally,
the Models 2611, 2612, 2635, 2636, and 4200-SMU will operate if they are programmed to the 20 V
range or lower.
For additional information on interlocks, see
(on page 2-5). Also see the reference
manuals for your SMUs.
Always perform tests with the Model 8010 lid closed. Performing tests with the test fixture
lid open may expose the user to electrical, burn, and projectile hazards.