scan.trigger.channel.stimulus
This attribute determines which trigger events cause the channel actions to occur.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Instrument reset
Scan reset
Create configuration
script
scan.trigger.EVENT_CHANNEL_READ
Y
Usage
eventID
= scan.trigger.channel.stimulus
scan.trigger.channel.stimulus =
eventID
eventID
Trigger stimulus used for the channel action; see
Details
for possible trigger
event IDs
Details
This attribute selects which events cause the channel event detector to enter the detected state. Set
this attribute to
0
to start the channel action immediately at the default setting.
Set
eventID
to one of the existing trigger event IDs shown in the following table.
Trigger event IDs
Trigger event ID
Description
channel.trigger[N].EVENT_ID or 41
to 48
The trigger event generated by the channel trigger
N
.
digio.trigger[N].EVENT_ID or 1 to
14
An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
display.trigger.EVENT_ID or 39
The trigger key (TRIG) on the front panel is pressed.
dmm.trigger.EVENT_LIMIT1_HIGH or
53
A DMM trigger event that indicates a measurement
has exceed the high limit value on limit 1.
dmm.trigger.EVENT_LIMIT1_LOW or
52
A DMM trigger event that indicates a measurement
has exceed the low limit value on limit 1.
dmm.trigger.EVENT_LIMIT2_HIGH or
55
A DMM trigger event that indicates a measurement
has exceed the high limit value on limit 2.
dmm.trigger.EVENT_LIMIT2_LOW or
54
A DMM trigger event that indicates a measurement
has exceed the low limit value on limit 2.
trigger.EVENT_ID or 40