Models 707B and 708B Switching Matrix Reference Manual
Section 7: TSP command reference
707B-901-01 Rev. B / January 2015
7-73
Usage
lockout
= display.locallockout
display.locallockout =
lockout
lockout
0
or
display.UNLOCK
: Unlocks EXIT (LOCAL) key
1
or
display.LOCK
: Locks out EXIT (LOCAL)
key
Details
Set
display.locallockout
to
display.LOCK
to prevent the user from interrupting remote
operation by pressing the EXIT (LOCAL) key.
Set this attribute to
display.UNLOCK
to allow the EXIT (LOCAL) key to interrupt script or remote
operation.
Example
display.locallockout = display.LOCK
Disables the front-panel EXIT (LOCAL) key.
Also see
None
display.menu()
This function presents a menu on the front panel display.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
selection
= display.menu(
name
,
items
)
selection
Name of the variable that holds the selected menu item
name
Menu name to display on the top line
items
Menu items to display on the bottom line