Model 6517B Electrometer User's Manual
Section 4: Basic measurements
6517B-900-01 Rev. B August 2022
4-27
External temperature
The Keithley Instruments Model 6517-TP temperature probe is a K-type thermocouple sensor that
measures external temperature. This sensor plugs into the TEMP TYPE K connector on the rear
panel of the 6517B. You can measure temperatures from –25 °C to 150 °C using the Model 6517-TP
temperature probe.
Do not allow the exposed thermocouple sensor to come in contact with chassis ground or to a voltage
potential. This causes erroneous readings. The thermocouple must be electrically insulated before
attaching it to a device that is at chassis ground or has a voltage on it.
To make temperature measurements:
1. Connect the 6517-TP temperature probe to the
TEMP TYPE K
connector on the rear panel.
2. On the front panel, press the
MENU
key.
3. Select the
GENERAL
menu. Press the
ENTER
key.
4. Select
A/D CONTROLS
. Press the
ENTER
key.
5. Select
DATA-STAMP
. Press the
ENTER
key.
6. Select
TEMPERATURE
.
7. Use the
RANGE
keys to toggle temperature readings to
ON
. Press the
ENTER
key.
8. Press the
EXIT
key as needed to return to the measurement display.
To change the temperature units:
1. On the front panel, press the
MENU
key.
2. Select the
GENERAL
menu. Press the
ENTER
key.
3. Select the
DISPLAY
menu. Press the
ENTER
key.
4. Select
TEMP-UNITS
. Press the
ENTER
key.
5. Set the units to
°C
,
°F
, or
K
. This setting also affects the display of internal temperature
measurements.
6. Press the
ENTER
key.
Internal scanning
You can use a 6517B with a scanner card, such as the Keithley Instruments Model 6521 or Model
6522, installed in the option slot. The following topics provide basic information for scanning internal
channels. If the scanner card is not already installed, refer to the scanner card instruction manual.
For complete information on scanning, refer to the scanner card instruction documentation. The
scanner card must be installed in the option slot to access the scan menus.
Содержание Keithley 6517B
Страница 2: ...Electrometer User s Manual Model 6517B ...