Section 4: Basic measurements
Model 6517B Electrometer User's Manual
4-26
6517B-900-01 Rev. B August 2022
Relative humidity and temperature readings
With the appropriate equipment connected, the 6517B can measure relative humidity and external
temperature. To view the humidity and external temperature readings on the front panel, select the
NEXT display. An example of the display is shown in the following figure.
Figure 43: Relative humidity and external temperature display
These readings can also be included as data elements for the bus interfaces (GPIB and RS-232) and
the buffer.
Relative humidity
The Keithley Instruments Model 6517-RH humidity probe measures relative humidity. This sensor
plugs into the HUMIDITY connector on the rear panel of the 6517B. You can measure relative
humidity from 0% to 100% using the Model 6517-RH.
To make relative humidity measurements:
1. Connect the 6517-RH humidity probe to the
HUMIDITY
connector on the rear panel.
2. On the front panel, press the
MENU
key.
3. Select the
GENERAL
menu. Press the
ENTER
key.
4. Select
A/D CONTROLS
. Press the
ENTER
key.
5. Select
DATA-STAMP
. Press the
ENTER
key.
6. Select
HUMIDITY
.
7. Use the
RANGE
keys to toggle relative humidity readings to
ON
. Press the
ENTER
key.
8. Press the
EXIT
key as needed to return to the measurement display.