Section 3: Setting up SMUs in Clarius
Model 4200A-SCS Source-Measure Unit (SMU) User's Manual
3-16
4200A-SMU-900-01 Rev. A December 2020
Current Step operation mode
The Current Step operation mode increments through evenly-spaced, constant current steps over a
range that you specify. The time interval for each step is determined automatically by the time
required to complete a sweep.
Figure 45: Stepping the diode current
For each step, parametric curve data is generated that is recorded in the Analyze pane.
Common operation mode - SMU
Common operation mode maintains a zero-voltage state at the terminal, subject to the maximum
current compliance of the connected SMU. The 4200-SMU and 4201-SMU each have a compliance
of 105 mA. The 4210-SMU and 4211-SMU each have a compliance of 1.05 A.
You cannot set any parameters for the Common operation mode.
SMU - all terminal parameters
When you select
All Parameters
, the Configure pane displays all available parameters for the
terminals for the test that is selected in the project tree.
Parameter descriptions are provided in the following topics.
Base Current
Available when Pulse Mode is selected. The current level that the instrument sources when the pulse
output is off. The level that you can set depends on the present source range.
Base Voltage
Available when Pulse Mode is selected. The voltage level that the instrument sources when the pulse
output is off. The level that you can set depends on the present source range.
Bias
The bias is the amount of voltage or current to be forced.