![Tektronix Keithley 4200-SMU Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/tektronix/keithley-4200-smu/keithley-4200-smu_user-manual_1078017011.webp)
Section 1: Introduction
Model 4200A-SCS Source-Measure Unit (SMU) User's Manual
1-2
4200A-SMU-900-01 Rev. A December 2020
The source of the SMU can be configured to sweep or step voltages or currents, or to output a
constant bias voltage or current.
There are medium-power and high-power source-measure units available. The 2 W medium-power
SMUs are models 4200-SMU and 4201-SMU. The 20 W high-power SMUs are models 4210-SMU
and 4211-SMU. The following table lists the maximum limits of the SMUs.
Source-measure units
Model
Maximum voltage
Maximum current
Maximum power
4200-SMU
4201-SMU
210 V
105 mA
2.2 W
4210-SMU
4211-SMU
210 V
1.05 A
22 W
Preamplifier
A 4200-PA preamplifier adds five low-current source-measure ranges to a SMU. Without a
preamplifier, the 100 nA range (100 fA resolution) is the lowest current source-measure range for a
SMU. With a preamplifier installed, the 10 nA, 1 nA, 100 pA, 10 pA, and 1 pA source-measure ranges
are added.
If preamplifiers are ordered, the 4200A-SCS is shipped from the factory with the preamplifiers
installed on the rear panel of the mainframe.
Ground unit (GNDU)
The ground unit on the rear panel of the 4200A-SCS provides a convenient method of making ground
connections. This eliminates the need to use a SMU for this purpose.