Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-177
:SYSTem:VERSion?
This command queries the present SCPI standard version.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:VERSion?
Details
This query command returns the SCPI standard version.
Example
:SYST:VERS?
Query the version. Example output:
1999.0
Also see
None
TRACe subsystem
The TRACe subsystem contains commands that control the reading buffer.
:DATA[n]:CLEar
This command clears all readings in the buffer.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:DATA[n]:CLEar
[n]
Channel number: For single-channel instruments, the value is
1
Example
:DATA:CLE
Clear the data buffer.
Also see
(on page 3-20)