Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 7: SCPI command reference
077114601 / March 2019
7-155
Example
:STAT:OPER:INST:ISUM?
Returns the contents of the Operation
Instrument Summary Event Register of the
status model.
Also see
Operation Instrument Summary Event Register
(on page D-13)
:STATus:OPERation:INSTrument:ISUMmary:ENABle
This command sets or reads the contents of the Operation Instrument Summary Enable Register of the status
model.
Type
Affected by
Where saved
Default value
Command and query
Power cycle
Not applicable
0
Usage
:STATus:OPERation:INSTrument:ISUMmary:ENABle <n>
:STATus:OPERation:INSTrument:ISUMmary:ENABle?
[n]
The status of the Operation Instrument Summary Enable Register
Details
This command sets or reads the contents of the enable register of the Operation Instrument
Summary Event Register (OISER).
When one of these bits is set and the corresponding bit in the OISER is set, bit 1 in the Operation
Instrument Event Register is set.
When sending binary values, preface
<n>
with
#b
. When sending hexadecimal values, preface
<n>
with
#h
. No preface is needed when sending decimal values.
Example
:STAT:OPER:INST:ISUM:ENAB #b0000000100010000
Sets the filter and measure bits of the
Operation Instrument Summary Enable
Register using a binary value.
You could also send the decimal value
272
or the hexadecimal value
#h110
Also see
Operation Instrument summary Event Register
(on page D-13)