Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-114
077114601 / March 2019
SENSe subsystem
The SENSe subsystem commands configure and control the measurement functions of the 2281S.
Many of these commands are set for a specific function (concurrent, current, or voltage). For example,
you can program a range setting for each function. The settings are saved with that function.
:SENSe[n]:FUNCtion
This command selects which type of measurement is active: Concurrent, current, or voltage.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
"CONC"
Usage
:SENSe[n]:FUNCtion <name>
:SENSe[n]:FUNCtion?
[n]
Channel number: For single-channel instruments, the value is
1
<name>
A string that contains the measurement function to make active:
Concurrent
: "CONCurrent[:DC]"
Current
: "CURRent[:DC]"
Voltage
: "VOLTage[:DC]"
(on page 2-96).
Details
Set this command to the type of measurement you want to make.
Reading this command returns the measure function that is presently active.
Examples
:SENS:FUNC "VOLT"
Make the voltage measurement function the active
function.
Also see
None