Section 7: SCPI command reference
Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
7-6
077114601 / March 2019
*OPC
This command sets the operation complete (OPC) bit after all pending commands, including overlapped
commands, have been executed.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Not applicable
Not applicable
Usage
*OPC
*OPC?
Details
When the
*OPC
command is sent, the OPC bit (bit 0) in the Status Event Status Register is set after
all pending command operations have been executed.
When the
*OPC?
command is sent, an ASCII "1" is placed in the output queue after all pending
command operations have been executed.
After all programmed operations are completed, the instrument returns to the idle state, at which time
all pending commands (including
*OPC
and
*OPC?
) are executed. After the last pending command is
executed, an ASCII "1" is placed in the output queue.
Also see
(on page 7-106)
*RCL
This command returns the instrument to the setup that was saved with the
*SAV
command.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
*RCL <n>
[n]
An integer from 1 to 5 that represents the saved setup
Details
Restores the state of the instrument from a copy of its settings that are stored in the setup memory.
The settings are saved using the
*SAV
command.
Example
*RCL 3
Restores the settings stored in memory location
3.
Also see
(on page 7-7)
(on page 2-105)