Model 2281S-20-6 DC Power Supply and Battery Simulator Reference Manual
Section 5: Application examples
077114601 / March 2019
5-7
To configure and execute a list sweep using SCPI commands:
*RST
:SENS:CONC:NPLC 1
:LIST1:VOLT "1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0, 9.0, 10.0"
:LIST1:CURR "2.0, 2.0, 2.0, 2.0, 2.0, 2.0, 2.0, 2.0, 2.0, 2.0"
:LIST1:DWEL "3.0, 3.0, 3.0, 3.0, 3.0, 3.0, 3.0, 3.0, 3.0, 3.0"
:LIST:SAVE:INTE 1
:LIST:RCL 1
:ARM:SOUR BUS
:TRIG:SOUR IMM
:TRIG:COUN 10
:TRIG:SAMP:COUN 10
:TRAC:CLE
:TRAC:FEED:CONT NEXT
:TRAC:FEED SENS
:TRAC:POIN 100
:LIST:STAT ON
:INIT
*TRG
*WAI
:TRAC:DATA? "READ,SOUR"
Perform a fast current load measurement
The following example demonstrates how to configure the power supply to perform a fast current
measurement initiated by an external trigger on digital line 1.
This measurement is made in the power supply function.
Equipment required
One 2281S power supply
Dual-channel arbitrary function generator (Tektronix AFG3102C)
Oscilloscope
Resistors: 2.5 Ω and 20 Ω
nMOSFET