Error Messages
Series 2268 850 Watt DC Power Supplies Reference Manual
246
2268S-901-01 Rev. A / January 2015
Device-Specific Error List
An error in the range [-399, 300] or [1, 32767] indicates that the instrument has detected an
error which is not a command error, a query error, or an execution error; some device
operations did not properly complete, possibly due to an abnormal hardware or firmware
condition. These codes are also used for self-test response errors. The occurrence of any error
in the class causes the device-specific error bit (bit 3) in the Event Status Register to be set.
Device-Specific Error List
Error
Code
Error Message Description
-300
Device-specific error. Indicates that the power supply could not
complete the operation due to some condition of the power
supply.
-313
Calibration Memory Lost
-314
Save/Recall Memory Lost
-315
Configuration Memory Lost
-350
Queue Overflow A specific code entered into the queue in lieu of
the code that caused the error. This code indicates that there is no
room in the queue and an error occurred but was not recorded.
-360
Communication error
This is the generic communication error for devices.
-363
Input buffer overrun Software or hardware input buffer on serial
port overflows with data caused by improper or nonexistent
pacing.
-365
Time out error
This is a generic device-dependent error.
Query Error List
An error number in the range [-499, -400] indicates that the output queue control of the
instrument has detected a problem with the message exchange protocol described in IEEE
488.2, section 6. The occurrence of any error in this class causes the query error bit (bit 2) in
the Event Status Register to be set.
Query Error List
Error
Code
Error Message Description
-400
This is the generic query error for the power supply, used only
when the other types of errors do not apply.