2260B Series User Manual
96
Exit
Press the Utility key again to exit
the Test sequence settings. The
Utility key light will turn off.
Load Test Sequence from USB
Overview
Before a test sequence can be run, it must first
be loaded into a one of the 10 memory save
slots. Before loading a test sequence into
memory:
Ensure the test sequence file is placed in the root
directory.
Ensure the file name number corresponds to the
memory number that you wish to save to.
For example: A test sequence file named
t001.csv can only be saved to memory
number 01, t002.csv can only be saved to
memory number 02, and so on.
Steps
1.
Insert a USB flash drive into the
front panel USB-A slot. Ensure the
flash drive contains a test sequence
file in the root directory.
2.
Turn on the power. MS (Mass Storage) will be
displayed on the screen after a few seconds if
the USB drive is recognized.