Remote programming
This appendix describes commands and queries that can be sent to the sensor head when attached to a Tektronix
oscilloscope. Long-form and short-form keywords are indicated with upper/lower case letters. The commands and queries
are supported by most oscilloscopes; differences in supporting oscilloscopes, if any, are described with the commands.
For additional information, refer to the programmer documentation for your oscilloscope.
Command list
A list of the commands supported by TIVP in alphabetical order.
CH<x>:PRObe? (Query Only)
This query-only command returns all information concerning the probe that is attached to the specified channel. The
channel is specified by x.
Syntax
CH<x>:PRObe?
Examples
CH2:PROBE?
might return
1.0000E-01; RESISTANCE 1.0000E
+07;UNITS "V";ID:TYPE "10X"'SERNUMBER "N/A"
for a 10X
probe, indicating that (among other parameters) the attenuation factor for the probe
attached to Channel 2 is 100.0 mV (assuming that probe units are set to volts).
CH<x>:PRObe:AUTOZero (No Query Form)
This command executes the AutoZero function. The operation is entirely performed by the oscilloscope. The channel is
specified by x.
Refer to the self-calibration procedure for information on performing the self calibration.
Syntax
CH<x>:PRObe:AUTOZero EXECute
Arguments
EXECute
sets the probe attached to the specified channel to AutoZero.
Examples
CH1:PROBE:AUTOZERO EXECUTE
sets the probe attached to the Channel 1
to autozero.