Reference information
Specifications
The following tables list the specifications for the measurement system. The specifications are Typical unless noted
otherwise.
Note: Preliminary specifications are indicated by an asterisk (*)
The performance limits in this specification are valid with these conditions:
• The instrument must be in an environment with temperature, altitude, and humidity within the operating limits described
in these specifications.
• The instrument must have had a warm-up period of at least 20 minutes.
• The measurement system is powered from a TekVPI compatible oscilloscope.
Warranted specifications describe guaranteed performance with tolerance limits or certain type-tested requirements.
The performance verification procedures are listed later in this document.
Performance verification procedures
on page 45
Overview
Characteristic
TIVP1
TIVP05
TIVP02
Bandwidth
1 GHz
500 MHz
200 MHz
Rise time
450ps
850ps
2ns
Differential Input Voltage Range,
Offset Range, Differential
Impedance
Use only the sensor tip cables listed.
Sensor tip cable
Differential input
voltage range
Offset range
Input impedance
SMA Input (50 Ω
mode)
±5 V
±25 V
50 Ω || N.A.
SMA Input (1 MΩ
mode)
±5 V
±25 V
1 MΩ || 11 pF
TIVPMX10X
±50 V
±200 V
10 MΩ || 2.8 pF
TIVPMX50X
±250 V
±250 V
10 MΩ || <5 pF*
TIVPSQ100X
±500 V
±500 V
10 MΩ || <5 pF*
TIVPWS500X
±2.5 kV
±2.5 kV
40 MΩ || <4 pF*
TIVPMX1X
±5 V
±25 V
50 Ω or 1 MΩ || 11
pF
Common Mode Rejection Ratio
Sensor tip
cable
DC
1 MHz
100 MHz 200 MHz 500 MHz 1 GHz
SMA Input
(50 Ω mode)
160 dB
145 dB
100 dB
100 dB
100 dB
90 dB
SMA Input (1
MΩ mode)
160 dB
145 dB
100 dB
100 dB
100 dB
90 dB
TIVPMX10X 160 dB
115 dB
92 dB
90 dB
85 dB
80 dB
TIVPMX50X 160 dB*
104 dB*
85 dB*
80 dB*
73 dB*
70 dB*
Table continued…