Getting started
Learn about the controls, indicators, and the basic operations of the probe.
Operating considerations
Read this section before installing your measurement system to be aware of the operating requirements and clearance
requirements including possible hazardous areas when the measurement system is connected to the DUT.
Measurement system handling best practices
The measurement system consists of quality parts and should be treated with care to avoid damage or degrading the
performance due to mishandling. Consider the following precautions when handling the fiber-optic cables and sensor tip
cables.
• Do not crush, crimp, or sharply bend the fiber-optic cable. Avoid making loops in the fiber-optic cable smaller than
5 inches (12.7 cm).
• Do not twist the fiber-optic cable; twisting the cable will stress the optical fibers.
• Do not allow kinks or knots to develop in the fiber-optic cable.
• Avoid putting tension on the fiber-optic cable.
• Do not pull or jerk the fiber-optic cable, especially when kinks or knots are present.
• Do not drop the sensor head or comp box assembly since damage and misalignment of the internal optical components
can result.
• Avoid over-bending the sensor tip cables; do not exceed the minimum bend radius of 2.0 inches (5.1 cm).
• Avoid crushing the cables by accidentally running over the cable with a chair wheel or by dropping a heavy object onto
the cable.
• Never support the weight of the sensor head or comp box by the fiber-optic cable.
• Store the measurement system in the supplied carrying case when not in use.
Getting started
TIVP Series IsoVuTM Measurement System User Manual
24