Theory of Operation
Each Li-Ion battery pack has an internal microprocessor that monitors its charge
level. To check the battery status, remove the battery from the analyzer. Then
press the
PUSH
button on the side of the battery and note the charge value on the
LED charge indicator.
Taskbar Power Status Icon.
A battery icon appears in the taskbar when the
instrument operates from battery power. The icon displays the battery charge level
in 10% increments. If two batteries are installed, the icon displays the average
charge level of the two combined batteries. When the instrument operates from
AC power (from the Desktop Power Supply) a plug icon appears in the Taskbar
status area.
Battery System.
The instrument uses up to two Inspired Energy type NI2020HD24
10.8 V, 7200 mAH, Li-ion batteries. You do not need special tools to replace the
batteries. (See page 4-24,
Lithium-Ion Battery Maintenance
Battery operation characteristics include:
The battery power system operates the platform for approximately 9 hrs (12
watts load) from two Li-Ion batteries and approximately 4.5 hrs from a single
battery. Run time for the platform plus application varies depending on the
applications being run.
Only one of two batteries will be in use at any given time when two batteries
are installed.
If two batteries are installed, you can remove and replace either battery
without disrupting service.
If a second battery is installed, the platform automatically switches operation
to the second battery when the
fi
rst battery reaches cut off capacity (5%
reserve) or is removed. The transition is transparent and does not interrupt
service.
Battery Power Management Utility.
The Batteries tab in the Power Management
dialog box provides general information on battery status. This is the most accurate
information to determine battery charge status and probable manufacture dates.
(See page 4-26,
H600 and SA2600 Instrument-Speci
Measurement Module
The Measurement module contains no user-servicable parts. There are also no user
adjustments on the Measurement module. The entire instrument must be returned
for service. Service for the measurement module is not independently available.
H600 & SA2600 Service Manual
2–9
Содержание H600 RFHawk
Страница 1: ...xx H600 RFHawk SA2600 Spectrum Analyzer ZZZ Service Manual P071264101 071 2641 01...
Страница 2: ......
Страница 3: ...H600 RFHawk SA2600 Spectrum Analyzer ZZZ Service Manual xx Revision A www tektronix com 071 2641 01...
Страница 6: ......
Страница 14: ...Service Safety Summary viii H600 SA2600 Service Manual...
Страница 17: ...Operating Information...
Страница 18: ......
Страница 41: ...Theory of Operation...
Страница 42: ......
Страница 52: ...Theory of Operation 2 10 H600 SA2600 Service Manual...
Страница 53: ...Adjustment Procedures...
Страница 54: ......
Страница 56: ...Adjustment Procedures 3 2 H600 SA2600 Service Manual...
Страница 57: ...Maintenance...
Страница 58: ......
Страница 65: ...Removal and Replacement Procedures Rear Case Remove Battery Compartment H600 SA2600 Service Manual 4 7...
Страница 69: ...Removal and Replacement Procedures H600 SA2600 Service Manual 4 11...
Страница 89: ...Replaceable Parts...
Страница 90: ......
Страница 92: ...Replaceable Parts Figure 5 1 Module and cover replaceable parts 5 2 H600 SA2600 Service Manual...
Страница 94: ...Replaceable Parts Figure 5 2 Mainframe replaceable parts 5 4 H600 SA2600 Service Manual...
Страница 96: ...Replaceable Parts Figure 5 3 H600 and SA2600 accessories 5 6 H600 SA2600 Service Manual...
Страница 98: ...Replaceable Parts 5 8 H600 SA2600 Service Manual...
Страница 99: ...Specifications...
Страница 100: ......