Command Groups
Table 2-6: Calculate commands (cont.)
Command
Description
Switches on/off the mathematical function.
Switches on/off the complete post-processing block.
Selects postprocessing for totalize.
Calibration Subsystem
This subsystem controls the calibration of the interpolators used to increase the
resolution of the instrument.
Table 2-7: Calibration commands
Command
Description
Sets or returns whether the instrument calibrates the
time interpolators for every measurement.
Con
fi
gure Function
The CONFigure command sets up the instrument to make the same measurements
as the MEASure query, but without initiating the measurement and fetching the
result. Use con
fi
gure when you want to change any parameters before making
the measurement.
CONFigure; READ?
The CONFigure command causes the instrument to choose an optimal setting for
the speci
fi
ed measurement. CONFigure may cause any device setting to change.
READ? starts the acquisition and returns the result.
This sequence operates in the same way as the MEASure command, but now it is
possible to insert commands between CONFigure and READ? to
fi
ne-tune the
setting of a particular function. For example, you can change the input impedance
from 1M
Ω
to 50
Ω
.
Start with the command CONFigure:FREQ 2E6,1, where 2E6 is the expected
value 1 is the required resolution (1Hz).
Then send INPut:IMPedance 50 to set input impedance to 50
Ω
.
Then send READ? to start the measurement and returns the result.
CONFigure;INITiate;FETCh?
The READ? command can be divided into the INITiate command, which starts
the measurement, and the FETCh? command, which requests the instrument to
return the measuring results to the controller.
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-13
Содержание FCA3000 Series
Страница 2: ......
Страница 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 135: ...Status and Events ...
Страница 136: ......
Страница 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 157: ...Programming Examples ...
Страница 158: ......
Страница 179: ...Appendices ...
Страница 180: ......
Страница 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Страница 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...