Source Impedance
DTG5078 & DTG5274 & DTG5334 Data Timing Generator User Manual 2
1-73
Source Impedance
The source impedance of the output module is displayed. The source impedance of
the DTGM21 output module can be switched between 23
Ω
and 50
Ω
. The source
impedance is switched by replacing a jumper on the DTGM21. Other models are
fixed at 50
Ω
.
When the DTG application launches, source impedance of the output module
installed in each slot is checked and displayed. No user settings are needed in the
Level window.
DTGM32
Output Voltage (V
OUT
)
–0.07
×
R
L
+ Vtt V
OUT
0.07
×
R
L
+ Vtt
–2.00
V
OUT
7.00
High Level (V
OH
)
–1.00 V to +2.47 V into 50
Ω
to GND
–1.94 V to +7.00 V into 1 M
Ω
to GND
The following relational expressions must be satisfied:
V
OH
7.00
V
OH
(7.00
×
R
L
+ 50
×
Vtt ) / ( R
L
+ 50 )
V
OH
R
L
/ 50
×
( 2.5 – 0.06
×
R
L
/ ( R
L
+ 50 )) + Vtt
( –2.00
×
R
L
+ 50
×
Vtt ) / ( R
L
+ 50 ) V
OH
V
OH
Vtt – R
L
/ 50
Low Level (V
OL
)
–2.00 V to +2.44 V into 50
Ω
to GND
–2.00 V to +6.94 V into 1 M
Ω
to GND
The following relational expressions must be satisfied:
V
OL
–2.00
V
OL
(50
×
Vtt – 4.5
×
R
L
) / ( R
L
+ 50 )
V
OL
Vtt – R
L
×
(0.02 + 2.5 / ( R
L
+ 50 )) + Vtt
( –2.00
×
R
L
+ 50
×
Vtt ) / ( R
L
+ 50 ) V
OL
V
OL
<
(( 2.5 – 0.06)
×
R
L
/ 50 ) + Vtt
Amplitude (V
OH
–
V
OL
)
0.03 V
p-p
to 1.25 V
p-p
into 50
Ω
to GND
0.06 V
p-p
to 2.50 V
p-p
into 1 M
Ω
to GND
The following relational expressions must be satisfied:
( V
OH
– V
OL
)
>
2
×
( Vtt – R
L
/ 50 – Offset )
( V
OH
– V
OL
)
>
2
×
((( R
L
×
( –2 ) + 50
×
Vtt ) / ( R
L
+ 50 ) – Offset )
( V
OH
– V
OL
)
>
2
×
(( 2.5
×
R
L
– 50
×
50
×
Vtt ) / ( 2
×
R
L
+ 50 ))
( V
OH
– V
OL
)
>
2
×
(( 7
×
R
L
– 50
×
Vtt ) / ( R
L
+ 50 ) – Offset )
Resolution
5 mV
Table 1-22:
Range of output level parameters
(cont.)
Item
Range
≤
≤
≤
≤
≤
≤
≤
≤
≥
≥
≥
≥
≥
Содержание DTG5078
Страница 4: ......
Страница 10: ...Table of Contents iv DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 19: ...Reference...
Страница 20: ......
Страница 24: ...Operating Modes 1 4 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 34: ...Data Structure 1 14 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 58: ...Pattern Edit 1 38 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 88: ...Timing Parameters 1 68 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 100: ...DC Output 1 80 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 120: ...Jitter Generation 1 100 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 126: ...DTG5000 Configuration Utility 1 106 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 146: ...Pulse Generator mode 1 126 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 150: ...Offline mode 1 130 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 176: ...Diagnostics and Calibration 1 156 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 177: ...Appendices...
Страница 178: ......
Страница 186: ...Appendix B System Recovery B 4 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 189: ...Index...
Страница 190: ......
Страница 196: ...Index Index 6 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...