Level Calibration
Level Calibration calibrates the levels of the output signals in the total circuitry that
contains the
DTG5000 Series
and output modules combined. This calibration does
not require any cables or measuring equipments.
When to Perform
The DTG5000 software automatically executes Level Calibration every time it
starts. If you have set up the
DTG5000 Series
for Master-Slave operation, an attempt
to start the master machine (that is, the slave machine is already operating) causes
Level Calibration to work on all the channels of the master and slave machines.
Output level is affected by environmental temperature. Thus, you should perform
Level Calibration in the following cases:
The temperature of the
DTG5000 Series
mainframe varies during operation.
Optimum accuracy is required when you output the signal.
Operating Procedure
You can perform the calibration only by clicking the Execute button on the screen.
Before starting the calibration
Power on the
DTG5000 Series
mainframe. Warm up the mainframe for 20 minutes
at the normal operating temperature.
Starting the calibration
1.
Select
Level Calibration...
in the System menu. Then, press the
SELECT
key. The Level Calibration dialog box appears.
Содержание DTG5078
Страница 4: ......
Страница 10: ...Table of Contents iv DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 19: ...Reference...
Страница 20: ......
Страница 24: ...Operating Modes 1 4 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 34: ...Data Structure 1 14 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 58: ...Pattern Edit 1 38 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 88: ...Timing Parameters 1 68 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 100: ...DC Output 1 80 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 120: ...Jitter Generation 1 100 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 126: ...DTG5000 Configuration Utility 1 106 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 146: ...Pulse Generator mode 1 126 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 150: ...Offline mode 1 130 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 176: ...Diagnostics and Calibration 1 156 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 177: ...Appendices...
Страница 178: ......
Страница 186: ...Appendix B System Recovery B 4 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...
Страница 189: ...Index...
Страница 190: ......
Страница 196: ...Index Index 6 DTG5078 DTG5274 DTG5334 Data Timing Generator User Manual 2...