Commands Listed in Alphabetical Order
2.13889173040E-008, -5.00000000000E-001, 2.13843094343E-008,
-3.74686717987E-001, 1.97710825489E-008, -3.74686717987E-001
,
indicating the points in Mask 3.
MASK:MASK<x>:POINTSPcnt
This command sets or queries the points (vertices) in the speci
fi
ed mask in
percentage coordinates. The mask is speci
fi
ed by <x>, which can be 1 through 8.
Any currently existing points in the mask are deleted. The upper leftmost point is
0,0, and the lower rightmost point is 100,100.
Each mask polygon can have a maximum of 50 points or vertices. If more than
50 points are speci
fi
ed, an error is generated and the extra points are ignored. If
any point is outside of 0-100, an error is generated, and the point is clipped to the
nearest boundary. The order of the pairs has no effect on the mask created.
NOTE.
If the speci
fi
ed mask is unde
fi
ned, a query will return 0,0.
Group
Mask
Syntax
MASK:MASK<x>:POINTSPcnt <NR3>,<NR3>[,<NR3>,<NR3>, ...]
MASK:MASK<x>:POINTSPcnt?
Arguments
NR3
is the pair of xy coordinates for a point (vertex) in the mask. You must
specify more than two pair of points, otherwise the mask is marked as unde
fi
ned,
and the points are ignored. Resolution of the NR3 value is to the nearest pixel.
1000 pixels horizontally implies 0.1% resolution; 400 pixels vertically implies
0.25% resolution. The default is no points in the mask.
Examples
MASK:MASK7:POINTSPCNT 20.4,10.5,90,10.5,50,80
de
fi
nes the points in
Mask 7.
MASK:MASK7:POINTSPCNT?
might return
MASK:MASK7:POINTSPCNT
20.4000,10.5000,90.000,10.500,50.000,80.000
as the percentage
coordinates for Mask 7.
MASK:SOUrce
This command sets or queries which source will be compared against the mask(s)
when counting is turned on. This command can optionally set the timebase
of the source. This is equivalent to selecting a waveform (and timebase) in the
Source section of the Mask Setup dialog box. If you do not specify the timebase,
the default is Main.
2-216
DSA8300 Programmer Manual
Содержание DSA8300
Страница 1: ...xx DSA8300 Digital Serial Analyzer ZZZ Programmer Manual P077057002 077 0570 02 ...
Страница 2: ......
Страница 3: ...DSA8300 Digital Serial Analyzer ZZZ Programmer Manual xx Revision B www tektronix com 077 0570 02 ...
Страница 8: ...Preface iv DSA8300 Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 404: ...Commands Listed in Alphabetical Order 2 388 DSA8300 Programmer Manual ...
Страница 405: ...Status and Events ...
Страница 406: ......
Страница 426: ...Status and Events 3 20 DSA8300 Programmer Manual ...
Страница 427: ...Examples ...
Страница 428: ......
Страница 431: ...Appendices ...
Страница 432: ......
Страница 433: ...Appendix A Character Set DSA8300 Programmer Manual A 1 ...
Страница 434: ...Appendix A Character Set A 2 DSA8300 Programmer Manual ...
Страница 446: ...Appendix C Factory Default Setup Values C 8 DSA8300 Programmer Manual ...