Application Examples
4.
Calculate the ampli
fi
er gain using the following
equations:
Gain = (output amplitude ÷ input amplitude) =
(3.155 V ÷ 130.0 mV) = 24.27
Gain (dB) = 20 x log(24.27) = 27.7 dB
Customizing Your Measurements
In this example, you want to verify that the
incoming signal to digital equipment meets its
speci
fi
cations. Speci
fi
cally, the transition time
from a low logic level (0.8 V) to a high logic
level (2.0 V) must be 10 ns or less.