Analyze Waveform Data
Amplitude Measurements (cont.)
Measurement
Description
Pk-Pk
The absolute difference between the maximum and minimum amplitude in the entire
waveform or gated region.
Amplitude
The high value less the low value measured over the entire waveform or gated region.
High
This value is used as 100% whenever high reference, mid reference, or low reference
values are needed, such as in fall time or rise time measurements. Calculate using either
the min/max or histogram method. The min/max method uses the maximum value found.
The histogram method uses the most common value found above the midpoint. This value is
measured over the entire waveform or gated region.
Low
This value is used as 0% whenever high reference, mid reference, or low reference values are
needed, such as in fall time or rise time measurements. Calculate using either the min/max
or histogram method. The min/max method uses the minimum value found. The histogram
method uses the most common value found below the midpoint. This value is measured
over the entire waveform or gated region.
Max
The most positive peak voltage. Max is measured over the entire waveform or gated region.
Min
The most negative peak voltage. Min is measured over the entire waveform or gated region.
160
DPO3000 Series Oscilloscopes User Manual
Содержание DPO3000 Series
Страница 1: ...DPO3000 Series Digital Phosphor Oscilloscopes ZZZ User Manual xx www tektronix com 071 2410 01 ...
Страница 4: ......
Страница 16: ...Preface xii DPO3000 Series Oscilloscopes User Manual ...
Страница 39: ...Installation 3 If necessary adjust your probe Repeat as needed DPO3000 Series Oscilloscopes User Manual 23 ...
Страница 50: ...Installation 2 Power off your oscilloscope 34 DPO3000 Series Oscilloscopes User Manual ...
Страница 54: ...Installation 6 Power on the oscilloscope 7 Push Utility 38 DPO3000 Series Oscilloscopes User Manual ...