Active probes
Active voltage probes often contain preamplifiers that provide high resistance and low capacitance at the probe tip. Some active
voltage probes contain differential amplifiers that provide high Common Mode Rejection Ratio (CMRR).
Active current probes often contain Hall-effect sensors that extend the probe bandwidth down to DC.
Many active probes, especially high frequency probes, can drive a 50 Ω instrument input.
Amplitude
This voltage measurement is the high value minus the low value measured over the entire waveform or gated region:
Amplitude = High – Low
Analog oscilloscope sweeps
Area
Area is a voltage-over-time measurement. The measurement is the area over the entire waveform or gated region expressed in
volt-seconds. The area above the common reference point is positive while the area below the common reference point is
negative.
ASin
Inserts the text ASIN( into the math expression.
ATan
Inserts the text ATAN( into the math expression.
Auto trigger mode
Auto trigger mode causes the instrument to trigger if a valid trigger event does not occur within a set amount of time.
Auto trigger mode is useful when probing DC signal levels such as power supplies or logic levels that are stuck high or low. It is
also useful when looking for activities on signal nodes such as oscillations when you first set up the instrument. The length of
time it waits for a trigger event depends on the time base setting.
B-Event scan setup
B-Event Scan is particularly useful for creating overlapped eye diagrams, and applies to all standard Pinpoint Trigger A- and B-
Events sequenced by Trig on nth Event in non-FastAcq mode.
The B-Event Scan initializes the Start Event value to 1 and the End Event value to 8. The Start Event must be less than or equal
to the End Event value. Advance After, advances the B-Event value after the given number of acquisitions have occurred
depending on the selected Advance method (Sequential, Random, and Toggle).
Oscilloscope reference
744
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
Содержание DPO2304SX
Страница 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Страница 2: ......
Страница 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Страница 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 83: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 51...
Страница 89: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 57...
Страница 100: ...About Tektronix oscilloscopes 68 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 234: ...Horizontal and acquisition setups 202 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 276: ...Mask testing setups 244 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 558: ...Copying and printing 526 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Страница 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 687: ...Figure A Figure B Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 655...
Страница 898: ...Oscilloscope reference 866 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...