AFG1000 Series Arbitrary Function Generator Programmer Manual
Command Descriptions
Arguments
LINear sets the sweep spacing to linear.
LOGarithmic sets the sweep spacing to logarithmic.
Returns
LIN|LOG
Examples
SOURce1:SWEep:SPACing LINear
sets the CH1 sweep spacing to linear.
[SOURce[1|2]]:SWEep:TIME
This command sets or queries the sweep time for the sweep for the specified
channel. The sweep time does not include hold time and return time. The setting
range is 1 ms to 500 s.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:SWEep:TIME
{<seconds>|MINimum|MAXimum}
[SOURce[1|2]]:SWEep:TIME?
Arguments
<seconds>::=<NRf>[<units>]
where:
<NRf> is the sweep time in seconds.
<units>::=[ns | μs | ms | s]
Returns
<seconds>
Examples
SOURce1:SWEep:TIME 100ms
sets the CH 1 sweep time to 100 ms.