Functional Test Procedures
Verify Optical Input
Channels
This veri
fi
cation is done without an input signal.
Equipment required
None
Prerequisites
At least one optical (80C00 series) sampling module must be
installed.
1.
Initialize the instrument:
Push the front-panel DEFAULT SETUP button.
2.
Set the Trigger System:
In the UI application toolbar, select
Free Run
from
the Trig menu.
3.
Select the channel to test:
Push the channel button for the channel to test. The
button lights and the screen shows the selected channel signal.
Figure 3: Channel button location
4.
Verify that the channel is operational:
Con
fi
rm that the following statements
are true.
A trace displays at about center screen.
The Waveform readout for the channel under test shows a value as follows:
80C01, 80C02, 80C04, 80C09, 80C11, and 80C11B: 1 mW
80C03: 100