Runt Search configuration menu
Use the Runt search to mark a waveform where a pulse crosses one threshold but fails to cross a second threshold before
recrossing the first threshold.
To create a new runt search:
1. Tap Add New... Search.
2. Set the Search Type to Runt.
3. Select the search Source.
4. Use the menu fields to set the search parameters.
To change the settings on an existing search, double-tap the search badge and make necessary changes.
Runt Search configuration menu fields and controls.
Field or control
Description
Display
Sets the display of the mark icons on or off. If you have multiple searches defined, the control
turns off just the marks for the selected search.
Stop Acquisition if Event
Found
Stops input acquisition when the search event occurs. Default is not enabled.
Search Type
Set to Runt.
Source
Lists the source channel or waveform to use to trigger or search. Types that require multiple
inputs will replace this control with a different source definition control.
Mark When
■
Occurs: A runt signal event occurs.
■
< Limit: A runt signal event occurs that has a pulse width less than the specified time limit.
■
> Limit: A runt signal event occurs that has a pulse width greater than the specified time
limit.
■
= Limit: A runt signal event occurs that has a pulse width that is equal to the specified time
limit (±5%).
■
≠ Limit: A runt signal event occurs that has a pulse width that does not equal (is greater
than or less than) the specified time limit (±5%).
Time Limit
Sets the time period condition to be met.
Only available when Mark When = < Limit, > Limit, = Limit, or != Limit.
Polarity
Sets the polarity of the pulse to detect (positive pulse only, negative pulse only, or a positive or
negative pulse).
Upper Threshold
Sets the upper threshold amplitude level through which the signal must pass to be considered a
valid transition.
Lower Threshold
Sets the lower threshold amplitude level through which the signal must pass to be considered a
valid transition.
Copy Trigger Settings to
Search
Sets the search criteria to match the current oscilloscope trigger settings.
Copy Search Settings to
Trigger
Sets the current oscilloscope trigger settings to match the search criteria.
Menus and dialog boxes
MSO54, MSO56, MSO58, MSO58LP, MSO64 Help
257
Содержание 6 series
Страница 24: ...Product documents and support 4 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 42: ...Options 22 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 54: ...Install your instrument 34 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 63: ...9 FlexChannel probe connectors Getting acquainted with your instrument MSO54 MSO56 MSO58 MSO58LP MSO64 Help 43 ...
Страница 84: ...Getting acquainted with your instrument 64 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 102: ...Configure the instrument 82 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 148: ...Advanced triggering 128 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 154: ...Zooming on waveforms 134 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 438: ...Waveform acquisition concepts 418 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 448: ...Waveform display concepts 428 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...
Страница 483: ...Where σ is the Phase Angle Degree Measurement algorithms MSO54 MSO56 MSO58 MSO58LP MSO64 Help 463 ...
Страница 518: ...Index 498 MSO54 MSO56 MSO58 MSO58LP MSO64 Help ...