Section 4—
5110
MAINTENANCE
This
section of the
manual contains information on
preparation
for
use and
reshipment, information for
performing preventive maintenance, troubleshooting, and
corrective maintenance for this
instrument.
PREPARATION
FOR
USE
AND
RESHIPMENT
The
following information provides detailed installa
tion
and operating
voltage instructions for the
os
cilloscope.
Do not
defeat the grounding connection. Any in
terruption
of the grounding connection can create
an
electric-shock
hazard. Before making external
connections
to
this instrument, always ground the
instrument
first by
connecting the power-cord to
a
proper
mating
power outlet.
The color coding
of the cord conductors may be in
accordance
with the following table.
DETAILED
OPERATING
POWER
INFORMATION
This instrument can be operated from either
a 120-volt
or
240-volt nominal line-voitage source, 48 to 440 hertz. In
addition,
three regulating ranges are
provided for each
nominal
line-voltage source.
Table
4-1
POWER CORD CONDUCTOR IDENTIFICATION
Conductor
Color
Alternate Color
Ungrounded
(Line)
Brown
Black
Grounded (Neutral)
Blue
White
Grounding
(Earthing) Green-Yellow
Green-Yellow
To prevent damage to the instrument, always check
the
line-voltage information recorded
on
the rear
panel
before applying
power to the instrument.
Power
Cord
Information
WARNING
J
This
instrument
is
intended to
be operated from a
single-phase
earth-referenced
power source
having
one
current-carrying
conductor (the
Neutral Con
ductor)
near earth potential. Operation
from power
sources
where both
current-carrying conductors
are
live
with respect to earth
(such as phase-to-
phase on a three-wire system) is not
recommended,
since
only the Line Conductor has over-current
(fuse) protection within the
instrument.
This instrument has a three-wire power cord with a
polarized
two-pole, three-terminal
plug for connec
tion
to
the power
source and safety-earth. The
safety-earth
terminal of the plug
is directly con
nected
to the instrument frame. For electric-shock
protection, insert
this plug
only in
a mating outlet
with a
safety-earth contact.
Line-Voltage
and Regulating-Range Selection
This
instrument
can be operated from either a 120-volt
ora
240-volt nominal line-voltage source
with
power-line
frequencies of 48 to
440 hertz. In addition, three operating
ranges can
be selected within each
nominal line-voltage
source.
This
permits the oscilloscope transformer to
operate
from 100-volt, 110-volt, 120-volt, 200-volt, 220-
volt,
and 240-volt sources. The nominal
voltage and
regulating range for which the instrument is currently set,
is marked on
the rear panel of the instrument. Before
connecting the oscilloscope
to line-voltage power, always
check
the rear panel to
see
if the voltage marked there
complies
with the expected line-voltage usage. If
the
voltage
marked
is proper, a change
of line voltage or
regulating
range will
not be necessary.
To
select
a different operating range, use the following
procedure to obtain correct instrument operation from
the
line
voltage available:
1.
Disconnect the instrument from the power source.
2.
Remove
the
bottom cabinet panel of the instrument
(see Cabinet Removal in this section) to gain access
to the
LV
Power Supply circuit board.
4-1
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......