Specification and Performance
Check-5110
Table
2-9 (cont)
LIST
OF TEST EQUIPMENT REQUIREMENTS FOR PERFORMANCE
CHECK
Description
Performance
Requirements
Application
Example
Coaxial
cable
Impedance,
50 Ω; length,
42 inch; connectors,
bnc.
Steps 8
through 13.
a. TEKTRONIX part
012-0057-01.
Coaxial
cable
Impedance,
50 Ω, length,
18 inch;
connectors, bnc.
Steps 12,13.
a. TEKTRONIX part
012-0076-00.
1X passive
probe
Compatible
with
5A-series
amplifiers
used
in the
Oscilloscope.
Step
14.
a.
TEKTRONIX P6101
Probe.
b.
TEKTRONIX
P6062B
Probe.
Termination
Impedance, 50 Ω;
accuracy
within
2%; connectors, bnc.
Steps
9, 11,12.
a. TEKTRONIX
part
011-0049-01.
Tee
connector
Connectors, bnc.
Steps
12, 13.
a. TEKTRONIX
part
103-0030-00.
Screwdriver
Length, 3-inch shaft; bit
size,
3/32 inch.
Step
1.
a. Xcelite
R3323.
PRELIMINARY
PROCEDURE FOR
PERFORMANCE
CHECK
1.
Ensure
that all
power switches are off.
2.
Check
the rear panel of the oscilloscope to
ensure
the
indicated
line voltage and the line voltage source are
the same.
INITIAL
CONTROL SETTINGS
Set
the
following controls during
warm-up time:
OSCILLOSCOPE
INTENSITY,
FOCUS Set for well-defined trace
and
normal
brightness.
3.
Ensure
that all
test equipment is suitably adapted to
the line voltage to be
applied.
4.
If applicable, install the TM
500-series
test equip
ment into the test equipment power module.
5.
Install
a
dual-trace vertical amplifier
unit
into the left
vertical compartment
of
the oscilloscope.
6.
Install
a time-base unit
into the horizontal compart
ment of the oscilloscope.
AMPLIFIER
PLUG-IN
Display
Position
CH 1
Volts/Div
CH
1
Cal
CH
1
Input coupling
Trigger
Mode
On
Centered
1
Fully
clockwise
de
CH 1
CH 1
7.
Connect
the equipment under
test and
the test
equipment
to a suitable line
voltage source. Turn all
equipment
on
and allow at least 20 minutes for the
equipment
to
stabilize.
NOTE
Titles
for external controls of the oscilloscope are
capitalized
in
this
procedure
(e.g. INTENSITY,
POWER).
TIME BASE PLUG-IN
Display
Position
Seconds/Div
Seconds/Div Cal
Swp
Mag
Triggering
Triggering Source
Chop
Centered
1
ms
Fully
clockwise
Off
+ Slope, Auto Trig,
ac Couple
Composite
2-6
REV
A JUL 1979
Содержание 5110
Страница 6: ...5110 Fig 1 1 5110 Oscilloscope 2134 01 ...
Страница 22: ......
Страница 34: ......
Страница 42: ......
Страница 83: ...5110 I43XII3 2134 15 ...
Страница 84: ... i 143 I132 2134 15 BLOCK DIAGRAM ...
Страница 85: ... t 2 E I 5110 S LNHNOdWOD BOVJHaiNI ...
Страница 88: ...J4 O3 INTERFACE J INTERFACE A NOV 1978 ...
Страница 92: ...AUXILIARY BOARD 1143 2134 18 AUXILIARY BOARD ...
Страница 95: ...DEFLECTION AMPLIFIERs ...
Страница 102: ...LV POWERSUPPLY CALIB COMPONENTS A 4 L V Power Supply circuit board 5110 ...
Страница 107: ...SIGNALS OUT OPTION 7 ONLY SI 10 OPT 7 213 26 REV A FE BI979 SIGNALS OUT COPTION T ONLY n ...
Страница 114: ......
Страница 115: ...5110 OSCILLOSCOPE ...
Страница 116: ...FIG 2 MAINFRAME p 5110 OSCILLOSCOPE ...
Страница 117: ...1 ...
Страница 121: ...o ...
Страница 122: ...S 1 5110 OSCILLOSCOPE ...
Страница 123: ...13 FIG 4 RACKMOUNT CABINET 5110 OSCILLOSCOPE ...
Страница 124: ......