Removal and replace procedures
on page 35 for lists and exploded views of replaceable modules.
Any module inside of the chassis that does not have a remove and replace procedure requires that the entire instrument be returned to
Tektronix Service Center for service.
Warning: Before performing this or any other procedure in this manual, read the safety summaries found at the beginning of
this manual. Also, to prevent possible injury to service personnel or damage to the instrument components, read
on page 15.
Before performing any procedure in this subsection, disconnect the power cord from the line voltage source. Failure to do so could
cause serious injury or death.
Before removing the rear chassis, order the 8 pieces of pre-cut aluminum tape (Tektronix part number 253-0582-XX). The existing
tape has to be removed from the chassis to access the interior and cannot be re-applied after removal. The aluminum tape is only
needed for instruments with a serial number below B020000 or below C040000.
Note: Read the cleaning procedure before disassembling the instrument for cleaning.
Required equipment
Most assemblies in this instrument can be removed with a T-10 or T8 Torx® screwdriver tip.
Table 3: Required equipment for removal and replacement
Item No.
Name
Description
1
Screwdriver handle
Accepts Torx-driver bits
2
T-10 Torx tip
Used for removing instrument screws. Torx-driver bit for T-10 size
screw heads
3
T-8 Torx tip
Used for removing instrument screws. Torx-driver bit for T-8 size
screw heads
4
3/16 inch open-end wrench
Used to remove nut posts
5
Needle-nosed pliers, small or medium size
Used to pull off aluminum shielding tape from rear case and flat
cable.
6
Proper antistatic work environment
To prevent electrostatic damage to components whenever you work
on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on
an antistatic mat.
Remove front-panel knobs
Use this procedure to remove and replace front-panel knob assemblies.
Before you begin
• To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded electrostatic prevention
wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
About this task
There are eight knob assemblies that can be removed from the front case.
Maintenance
5 Series MSO Service Manual MSO54, MSO56, MSO58