Trigger Setup
The part of the record that occurs before the trigger is the pretrigger portion. The part that occurs after the trigger is the posttrigger
portion. Pretrigger data can help you troubleshoot. For example, to
fi
nd the cause of an unwanted glitch in your test circuit, you can
trigger on the glitch and make the pretrigger period large enough to capture data before the glitch. By analyzing what happens
before the glitch, you may uncover information that helps you
fi
nd the source of the glitch. Alternatively, to see what is happening in
your system as a result of the trigger event, make the posttrigger period large enough to capture data after the trigger.
Slope and Level
The slope control determines whether the
instrument
fi
nds the trigger point on the rising
or the falling edge of a signal.
The level control determines where on that
edge the trigger point occurs.
The oscilloscope provides a long horizontal bar
or bars across the graticule to temporarily show
the trigger level.
122
Tektronix 4000 Series Oscilloscopes User Manual
Содержание 4000 Series
Страница 1: ...Tektronix 4000 Series Digital Phosphor Oscilloscopes User Manual www tektronix com 071 2121 01 ...
Страница 4: ......
Страница 18: ...Preface xiv Tektronix 4000 Series Oscilloscopes User Manual ...
Страница 53: ...Installation 2 Power off your oscilloscope Tektronix 4000 Series Oscilloscopes User Manual 35 ...
Страница 57: ...Installation 6 Power on the oscilloscope 7 Push Utility Tektronix 4000 Series Oscilloscopes User Manual 39 ...
Страница 68: ...Installation 50 Tektronix 4000 Series Oscilloscopes User Manual ...
Страница 312: ...Application Examples 294 Tektronix 4000 Series Oscilloscopes User Manual ...