TABLE 6-1
Power-Supply Ripple and Regulation
Calibration— Type 323
Power
Supply
Maximum
Ripple
Maximum output
change o f supply
( re g u la tio n )
-5 -V o lt
10 millivolts
±0 .0 3 volt
+ 5-Volt
10 millivolts
±0.025 volt
+ 14-Volt
150 millivolts
± 2 .8 volts
+ 100-Volt
150 millivolts
± 5 .0 volts
+ 175-Volt
500 millivolts
± 1 4 volts
b. CHECK— Precision DC voltmeter for power-supply out
put change within the limits given in Table 6-1 while vary
ing the DC power source as given above. Power-supply
test points are shown in Fig. 6-4A.
c. Disconnect the precision DC voltmeter.
d. Turn off the POWER switch temporarily and replace
the high-voltage cover. Replacement instructions are given
in the Maintenance section of this manual.
* 9 3
Lower Defl
Plate DC Level
R40
VAR V /D IV
BAL
'Centering
§ 1 Plate DC Level
9. Check Low Batteries Indicator
a. Change the following control settings:
Vertical POSITION
TIME/DIV
INTENSITY
POWER
Midrange
1 ms
Normal intensity
ON
b. Slowly decrease the output voltage of the variable
DC power supply.
c. CHECK— LOW BATT light begins to flash when vari
able DC power supply output voltage is -f-6.25 volts, ±0.31
volt (use the precision DC voltmeter to monitor the applied
DC voltage at the EXT DC POWER jacks).
d. Return variable DC power supply output voltage to
+ 8 volts.
NOTE
If the internal batteries o f the Type 323 are
charged, the remainder o f this procedure can be
performed with battery powered operation.
Fig. 6 -5 . Location o f vertical adjustm ents (M a in b o a rd ).
11. Adjust Vertical X lO Balance
O
a. Return the VARIABLE VOLTS/DIV control to CAL.
b. Position the trace to the center horizontal line with
the vertical POSITION control.
c. Pull the X lO VERT GAIN switch out. (Note: to pre
vent changing knob position, the X lO VERT GAIN switch
can be actuated using the vertical POSITION control bracket
behind the front panel).
d. CHECK—Trace shift less than 1.5 division vertically.
e. ADJUST— VERT X lO BAL adjustment, R39 (see Fig.
6-5), to return the trace to the center horizontal line. Repeat
parts b through e for minimum trace shift as the X lO VERT
GAIN switch is pulled out.
f. Recheck step 10. If readjustment is necessary, recheck
this step also.
10.
Adjust Variable Volts/Division Balance
O
a. Position the trace to the center horizontal line with
the vertical POSITION control.
b. CHECK— Rotate the VARIABLE VOLTS/DIV control
throughout its range. Trace should not move more than
one division vertically (if the trace is not visible at all, preset
the VAR V/DIV BAL adjustment to bring the trace on screen).
c. ADJUST—VAR V/DIV BAL adjustment, R40 (see Fig.
6-5), for no trace shift as the VARIABLE VOLTS/DIV con
trol is rotated. If necessary, use the vertical POSITION
control to keep the trace on screen during this adjustment.
12.
Adjust Deflection Plate DC Level
O
a. Push the X lO VERT GAIN switch in.
b. Connect the precision DC voltmeter from the collector
of Q163 (case of Q163; see Fig. 6-5) to chassis ground.
c. Rotate the vertical POSITION control fully clockwise
and note the meter reading. Then rotate the vertical POSI
TION control fully counterclockwise and again note the
meter reading.
d. CHECK— Voltage range measured in part c centered
around + 5 0 volts.
6-9
Содержание 323
Страница 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Страница 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Страница 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Страница 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Страница 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Страница 72: ...Maintenance Type 323 4 16 Fig 4 14 M ain circuit board p a rtia l loft side vertical circuit components ...
Страница 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Страница 157: ...BL OCK DIAGRAM ...
Страница 158: ......
Страница 161: ...A TYPE 323 OSCILLOSCOPE ...
Страница 162: ...1 TYPE 323 OSCILLOSCOPE ...
Страница 163: ...FIG 2 CABINET ...
Страница 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...