Performance Check Procedure— 2236 Service
TRIGGER
Equipment Required (see Table 4-1):
Leveled Sine-Wave Generator (Item 2)
50-0 BNC Cable (Item 8)
50-0 BNC Termination (Item 9)
INITIAL CONTROL SETTINGS
c. Set the generator to produce a 10-MHz, 3.5-division
display.
Vertical (Both Channels)
POSITION
Midrange
VERTICAL MODE
CH 1
BW LIMIT
Off (button out)
VOLTS/DIV
50 mV
VOLTS/DIV Variable
CAL detent
INVERT
Off (button out)
Input Coupling
DC
Horizontal
POSITION
Midrange
HORIZONTAL MODE
A
A and B SEC/DIV
0.1 fis
SEC/DIV Variable
CAL detent
X I0 Magnifier
Off (knob in)
B DELAY TIME
POSITION
Fully counterclockwise
B TRIGGER
SLOPE
OUT
LEVEL
Midrange
A TRIGGER
VAR HOLDOFF
NORM
Mode
NORM
SLOPE
OUT
LEVEL
Midrange
A&B INT
CH 1
A SOURCE
INT
A EXT COUPLING
DC
1. Check Internal Triggering
a. Select the FREQ function.
b. Connect the leveled sine-wave generator output via a
50-0 cable and a 50-0 termination to the CH 1 OR X input
connector.
d. Set the CH 1 VOLTS/DIV switch to 0.5 V.
e. CHECK— Stable display can be obtained by adjusting
the A TRIGGER LEVEL control for each switch combination
given in Table 4-5.
Table 4-5
Switch Combinations for A Triggering Checks
A TRIGGER Mode
A TRIGGER SLOPE
NORM
OUT
NORM
IN
P-P AUTO
IN
P-P AUTO
OUT
f. Set the HORIZONTAL MODE switch to B.
g. CHECK— Stable display can be obtained by adjusting
the B TRIGGER LEVEL control in a position other than the
B RUNS AFTER DLY position for both the OUT and IN
positions of the B TRIGGER SLOPE switch.
h. Increase the generator output voltage to produce a
0.5-division display.
i. CHECK—The readout is stable and accurately dis
plays the input signal frequency.
j. Set the HORIZONTAL MODE switch to A.
4-12
Содержание 2236
Страница 10: ...2236 Service viii The 2236 Oscilloscope ...
Страница 28: ...Specification 2236 Service 1 18 ADD JAN 1985 Figure 1 2 Physical dimensions of the 2236 Oscilloscope ...
Страница 74: ...Theory of Operation 2236 Service 3 30 ...
Страница 75: ...Theory of Operation 2236 Service Figure 3 13 Typical waveforms for the Delta Time function with Valt HI 3 31 ...
Страница 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Страница 240: ...2236 Service CTM INTERFACE SIGNALS F ig u re 9 4 O s c illo s c o p e b a sic b lo c k diag ram 4294 46 ...
Страница 241: ... VERT DEFL 2236 Service 4 2 0 6 3 5 4 2 0 4 4 7 F ig u re 9 5 O scillo sco p e d e ta ile d b lo c k diag ram ...
Страница 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Страница 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Страница 253: ...A10 COUNTER TIMER MULTIMETER BOARD F ig u re 9 8 C irc u it b o a rd lo c a tio n Illu s tra tio n ...
Страница 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Страница 255: ...2 2 3 6 S e rv ic e A ll c o m p o n e n ts on th e A 2 A tte n u a to r b o a rd are lo c a te d on D ia g r a m ...
Страница 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Страница 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Страница 275: ......
Страница 280: ......
Страница 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Страница 311: ...1 2 3 4 5 6 7 8 9 A I B L c T D T E t F ___ ____G____ ___ H___ j ____ J _j K j_ I ____i_ i___ N _ _ j___ P j S ...
Страница 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Страница 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Страница 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Страница 323: ...9 ...
Страница 324: ...2 REV APR 1986 ...
Страница 325: ...2236 SERVICE ...
Страница 326: ......
Страница 332: ......
Страница 333: ...2236 SERVICE ...
Страница 334: ...2 4 2 1 ...
Страница 338: ......