Betriebsanleitung/Operating manual
E3905 / P3394
36
BD_BE 763 b
www.tecsis.de
If logger memory contains recordings already:
When key: ´store/Quit´ is pressed for 2 seconds, the choice for clearing the logger memory will be
displayed:
The selection can be made by key: ´
/Min´ and key: ´
/Max´. Key:
´store/Quit´ enters the choice.
If the logger memory is full, the display will show:
Viewing recorded measurings
Within the „LoGG Stor“ function the measurings can be viewed directly in the display not only by
means of a computer (like at „Func CYCL“): press 2 seconds ´set/Menu´. The first menu displayed
now is „rEAd LoGG“ (read logger data). After pressing
key: ´
/Tara´ the measurement recorded
last will be displayed, changing between the different values referring to the measurement also is
done by pressing key: ´
/Tara´.
Changing the measurement is done by pressing the key: ´
/Min´or key: ´
/Max´.
6.2
„Func-CYCL“: Automatic recording with selectable cycle time
The Logger-Cycle-Time is settable (p.r.t. Configuration).
For example „CYCL“ = 60: A measuring is recorded after each 60 seconds.
When the slow measurement "rAtE-Slo" is chosen, additionally a low power function is available:
„Lo.Po“. If „Lo.Po“ is on, the device only will take a measurement at the point of time of the
recording. In between the recordings the measuring shuts down. This decreases the power
consumption enormously and therefore is recommended e.g. for long time recordings where no
mains adapter is available.
Max. number of measurings: 1800 or resp. 5400.
Cycle time:
1...3600 seconds (=1h), selectable in
the configuration
A measuring contains of:
-
slow measurings (rAtE SLo):
-
Sensor 1 current value at point of storage
-
Sensor 1 Min.-Peak, Max-Peak since last storage
-
Sensor 2 current value at point of storage
-
Sensor 2 Min.-Peak, Max-Peak since last storage
-
Current difference of Sensor 1 – Sensor 2 at point of storage
-
Min.-Peak, Max-Peak of difference of Sensor 1 – Sensor 2 Min.-Peak,
Max-Peak since last storage
clear nothing
(cancel menu)
clear the last
recording
clear all
recordings