Operating manual 
E3907 

10 

BE 984 i 

www.tecsis.de 

 
 

Call menu "POINTDIS", after pressing ENTER  the final value of the sensor can be entered, by 

PLUS

 or 

MINUS

 the numeric character can be edited, by ENTER the next figure can be edited, if all 

figures are edited, by 

PLUS

  and/or 

MINUS

  the comma will be shifted. For the completion of the 

input press ENTER, thereafter the final value is being logged in the E3907 

UNIT 

  

Input of the physical unit, e.g.  Kg, Ncm, t, gr, kN, N•m, bar... 

DESIG   

Name of the sensor e.g. sensor 1,2,3, silo, tank, mixer, balance1, motor, test1.. 

CALIB    

Selection of the sensor type (active, passive, 4-20mA, with or without calib. control..) 

 
 

4.4 Operating example 

 
- The E3907 is switched on by pressing ENTER 

 

- Control time and date in SYSTEM menu or adjust if necessary 

 

- Adjust language 
Call menu 2 SYSTEM – 2.1 LANG. When ENTER button was pressed, the 3907 is in the selection mode. With keys 

PLUS

 and/or 

MINUS 

the language can be selected. With ENTER the selected language will be stored. With ESC 

the selected language will be discarded. 
 
- Adjust time 
Call menu 2 SYSTEM – 2.8 DATE. When ENTER button was pressed TIME will appear on the LCD. Press ENTER 
again and the time can be adjusted. With 

PLUS

 and 

MINUS 

the hours can be adjusted. After ENTER the minutes 

can be adjusted with 

PLUS

 and 

MINUS

. With ENTER the selected time will be stored. With ESC the selected time 

will be discarded 
 
- Change to the measuring mode 
From menu 1 MEASUR by pressing ENTER  the list of all sensors is  called. With 

PLUS

  and/or. 

MINUS

  a sensor 

parameter set is being chosen. With ENTER the GM 80 is being adjusted to this sensor parameter set. With ESC it 
is possible to switch back to menu 1 MEASUR from any mode. 
 
- Change measuring rate by measuring mode 
The measuring mode can be left by pressing ESC. After pressing ESC  again, the GM 80 will be in menu 1 
MEASUR. From there the menu 2 SYSTEM – 2.4 Rate can be called. Select RATE by key 

PLUS

 and/or 

INUS

. By 

pressing key ENTER the set measuring rate is being displayed first. By key 

PLUS

 and/or 

MINUS

 a new measuring 

rate can be adjusted. By ENTER the new measuring rate will be taken over, by ESC the new measuring rate will be 
discarded.. 
 
- Switch off E3907 
E3907 is switched off, when ESC button is pressed >3 seconds.  
 
 

4.5 Measuring with the E3907 
 

4.5.1 Selected sensor 

 
Force sensor with measuring range:  

200KN 

Sensitivity 

  

 

 

2mV/V 

Designation e.g.: 

 

 

Press 

Parameter set: Sensor no.: 

 

3  

Option:  

 

 

 

100% Calibration control 

 
 
 
 
 
 

Содержание E3907

Страница 1: ...Force Pressure Temperature Switch Service Operating manual E3907 Universal handheld measuring device with datalogger BD 984 i 02 2013...

Страница 2: ...y 63073 Offenbach am Main Valid for E3907 handheld measuring device with datalogger Copyright 2006 2015 tecsis GmbH Offenbach Germany Reproduction prohibited Reproduction in whole or in part only with...

Страница 3: ...interface 11 4 6 1 Transfer rate specification RS232 V 24 USB 11 4 6 2 Operation via the serial interface 11 4 6 3 Handling with serial interface USB 14 4 6 4 Operation and function principle of the E...

Страница 4: ...Operating manual E3907 BE 984 i www tecsis de...

Страница 5: ...arantee Errors and changes are reserved Intended Purpose Improper Usage A tecsis device is used for displaying processing and controlling or regulation of processes It shall not be used as the only to...

Страница 6: ...take responsibility for that residual risks and to keep them as low as possible The relevant regulations and laws must be observed Residual risks or dangers associated with the measurement device shou...

Страница 7: ...s The device may only be used by qualified personnel according to the technical data in connection with the special safety rules and regulations 1 8 Technical changes The product may not be changed co...

Страница 8: ...An assignment of sensors to the corresponding parameter set is documented on the rear of the device 3 2 Power supply Mains operation Power supply 6V DC min 1 8A Fuse Self reset able fuse 1 5A Under Ov...

Страница 9: ...evice from moisture condensation rain snow Protect the device from direct sunlight Protect the device from dust and dirt Protect the device from high or excessive ambient temperature Protect the devic...

Страница 10: ...nu ESC Switch off the device Exit one step back in the menu Scroll up Scroll down USB USB Socket for USB Connector Type B RS232 RS232 Socket for 3 5mm jack 3pol front GND middle RxD outside TxD 6V DC...

Страница 11: ...Operating manual E3907 www tecsis de BE 984 i 7...

Страница 12: ...Operating manual E3907 8 BE 984 i www tecsis de 4 2 Switching on the device Press the ENTER button to switch on the device 4 3 Menu description 4 3 1 Brief handling overview E3907...

Страница 13: ...w many measurements an average value is accumulated Applicable at e g vibrations control oscillations 2 5 PASS Password query After entering the password 9373 it is possible to enter menu 3 CALIB Ther...

Страница 14: ...appear on the LCD Press ENTER again and the time can be adjusted With PLUS and MINUS the hours can be adjusted After ENTER the minutes can be adjusted with PLUS and MINUS With ENTER the selected time...

Страница 15: ...uring every keystroke on ENTER while measuring For data logging adjust operating mode in the logger mode e g choose AUTO and interval time e g 10 sec and change to measuring mode A measured value with...

Страница 16: ...Out datalogger B 0x42 Delete datalogger C 0x43 Read Out current sensor parameter D 0x44 Read Out status E 0x45 Read Out complete status g 0x67 Change protocol setup k 0x6B ENTER l 0x6C PLUS m 0x6D MI...

Страница 17: ...mA and 100 load mA Dec point 0xXXXX XXXX 000 _5000___ 001 _5 000__ 010 _50 00__ 011 _500 0__ 100 5 000___ 0 load 2Byte HEX value MSB LSB 100 load 2Byte HEX value MSB LSB 4 6 2 3 Read out status Status...

Страница 18: ...ol status 1Byte 0xXXXX XXXX binary coded 1 no closing character is sent 1 send CR LF 1 send CR 1 send LF 4 6 2 12 Calculation of the Checksum CS and the Weighted Checksum wCS The calculation takes pla...

Страница 19: ...starting time of the measurement the current sensor designation the final value of the measuring range the adjusted measuring rate and the logger mode e g AUTO 1ms are saved in the logger NOTE All pr...

Страница 20: ...ress key or and choose between 0 Load or Sensitivity 0 Load Relieve sensor from load or Sensitivity Enter the value fort the sensitivity in V Press Key or and choose between 100 Load or Sensitivity 10...

Страница 21: ...0mA short circuit protected Bridge resistance min 350 Ohm only full bride 4 or 6 conductor Sensitivity up to 3 3mV V other values on request 100 control via E3907 applied via control input 4 8 2 Conne...

Страница 22: ...measures before installation Attention The information for the voltage supply in section 5 2 must be respected Cable connections Attention Never connect voltage supply on non used pins 5 3 Normal ope...

Страница 23: ...k the plug connections Repair Notation The products contain no parts that need to be or that can be serviced by the user himself Repairs may only be carried out by the tecsis GmbH If it can be assumed...

Страница 24: ...tecsis GmbH Carl Legien Stra e 40 44 63073 Offenbach am Main Germany Phone 49 69 5806 0 Fax 49 69 5806 7788 E Mail force tecsis de Internet www tecsis de...

Отзывы: