Z.TRAC (Zero Tracking Sensitivity)
⊗
Zero tracking allows the display to adjust for minor changes in the zero balance
of the scale. When enabled, the instrument will track weight readings within the
zero 'dead' band back to exactly zero at a maximum rate of 0.5 (SLOW) or 10
(FAST) graduations per second.
•
Options are: OFF, SLOW, FAST
•
Default: OFF
Z.RANGE (Allowable Zero Operating Range)
⊗
This setting restricts the range over which the Zero functions can operate.
•
Options are: -2_ 2, -1_ 3, -20_20, FULL
•
Default: -2_ 2 (–2% to +2%)
Z.BAND (Zero 'Dead' Band)
⊗
This is an adjustable margin either side of true zero that defines the zero 'dead'
band. The zero 'dead' band is used by the automated functions to determine
zero load (eg. a setting of 4 specifies that readings between –4.5 and 4.5 are
considered to be zero).
When the displayed weight reading is within this band the instrument displays
the zero band annunciator. Refer to Status Annunciators page 20.
•
Settable over the full weight range. Always enter a number in multiples
of display units. Refer to RES (Count-by Resolution)
page 31 for
more information.
•
Default: 0 (ie. –0.5 to 0.5 graduations)
R.ENTRY (Entry Means to Full Setup)
This setting can be used to restrict access to the full setup. The first entry to the
full setup menu using the push button on the rear of the instrument causes all
future access to the full setup to be restricted to the push button.
•
Options are: OFF, ON
•
Default: OFF
7.4.3. CAL (Scale Calibration)
Items within this group perform various calibration routines. For detailed scale
calibration procedures refer to Calibration page 42. Certain items in the Scale Build
can affect the calibration of the scale. Always check that these sections are
correctly configured to suit the current application before attempting to calibrate the
scale.
ZERO (Zero Calibration)
⊗
Select to perform Zero Calibration. While the zeroing is in progress the display
will show Z.in P. Refer to ZERO (Zero Calibration Routine) page 43.
SPAN (Span Calibration)
⊗
Select to perform Span Calibration. While the span calculation is in progress the
display will show S.in P. Refer to SPAN (Span Calibration Routine) page 43.