background image

OPERATING INSTRUCTIONS 

 

 CL 326

 

DIGITAL THERMOMETER

 

 
 
 

 

INTRODUCTION 

This instrument is a 3½

 

digit, compact-sized portable 

digital thermometer designed to use external K/J/T/E type 
thermocouples as temperature sensor. Temperature indica-
tion follows Reference Temperature/Voltage Tables 
(N.I.S.T. Monograph 175 Revised to ITS-90) for K/J/T/E 
type thermocouples.    
 

SAFETY INFORMATION 

It is recommended that you read the safety and opera-

tion instructions before using the thermometer. 
 

WARNING 

To avoid electrical shock, do not use this instrument 

when working voltages at the measurement surface over 

24V AC or DC. 

 

WARNING 

To avoid damage or burns, do not make temperature 

measurement in microwave ovens. 

 

CAUTION 

Repeated sharp flexing can break the thermocouple 

leads. To prolong lead life, avoid sharp bends in the 

leads, especially near the connector. 

 

SPECIFICATIONS 

ELECTRICAL 

Temperature Scale:

 Celsius or Fahrenheit user-selectable. 

Measurement Range: 

Thermocouple Range 
K-TYPE (0.1°C) 

-200°C to 1372°C, 

 

 -328°F to 1999°F 

J-TYPE (0.1°C)    -210°C to 1200°C, 
 

 -346°F to 1999°F 

T-TYPE (0.1°C) 

-200°C to 400°C, 

 

 -328°F to 752°F 

E-TYPE (0.1°C) 

-220°C to 1000°C, 

 

 -364°F to 1832°F 

Calibration Range: 

-210°C to 1372°C, (-364°F to 1999°F) 

Auto range: 

0.1°C /1°C , 0.1°F /1°F 

Accuracy: 

Accuracy is specified for operating temperatures over 
the range of 18°C to 28°C (64°F to 82°F), for 1 year, not 
including thermocouple error. 
 

±(0.1% rdg + 1°C) on -60°C to 1372°C 

 

±(0.1% rdg + 2°C) on -60°C to -220°C 

 

±(0.1% rdg + 2°F) on -76°F to 1999°F 

 

±(0.1% rdg + 4°F) on -76°F to -364°F 

ENVIRONMENTAL 

Ambient Operating Ranges: 

0°C to 50°C (32°F to 122°F) <80% R.H. 

Storage Temperature: 

-20°C to 60°C (-4°F to 140°F) <70% R.H. 
 

GENERAL 

Display:

 

3½ digit liquid crystal display (LCD) with a maximum 
reading of 1999. 

Polarity: 

Automatic,positive implied,negative polarity indication. 

Overrange: 

-OL is displayed. 

Zero: 

Automatic. 

Reading Rate: 

one time per second. 

Low battery indication: 

The “

” is displayed when the battery voltage drops 

below the operating level. 

Measurement rate

1 times/second. 

Accuracy: 

Stated accuracy at 23°C±5°C, <75% R.H. 

Input Connector: 

Accepts standard miniature thermocouple connectors 
(flat blades spaced 7.9mm, center to center). 

Battery Life: 

100 hours typical with carbon zinc battery. 

Auto power off: 

The meter key switch inactive for more than 70 minutes. 

Temperature Coefficient: 

0.1 times the applicable accuracy specification per °C 
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 
82°F to 122°F). 

Input Protection: 

24V dc or 24V ac rms maximum input voltage on any 
combination of input pins. 

Maximum Differential Common Mode Voltage 
(Maximum Voltage between T1 and T2 during mea-
surement): 

1volt. 

Dimensions: 

195mm(H) x 92mm(W) x 53mm(D). 

Weight: 

approx. 9 oz. (250g) including battery.

 

°C /°F Selecting the Temperature Scale 

Readings is displayed in either degrees Celsius(°C) or 
degrees Fahrenheit(°F). 
When the thermometer is turned on, it is set to the tem-
perature scale that was in use when the thermometer 
was last turned off. 
To change the temperature scale, press the °C or °F key. 

 

” Button

 

Press “

” button to toggle on and off of backlight. 

The backlight will switch-off automatically after 30 
seconds. 

 

HOLD Mode   

Pressing the HOLD key to enter the Data Hold mode, 
the “HOLD” annunciator is displayed. 
When HOLD mode is selected, the thermometer held 
the present readings and stops all further measurements. 
Pressing the HOLD key again to cancel HOLD mode 
causing thermometer to resume taking measurements. 
 

K/J/T/E Input Thermocouple Type 

The TYPE key switch the input circulating selects the 
K/J/T/E type thermocouple as input. 
 

T1 T2/T1-T2 Input Selection 

The input selection indicates which input is selected; T1 
thermocouple, T2 thermocouple or the difference 
between the two thermocouples (T1-T2). 
When the thermometer is turned on, it is set to the tem-
perature input that was in use when the thermometer 
was last turned off. 