Check that the C2 lens (LC “Intensity”) is not severely under- or overfocused,
spreading the beam to a point that it is too dim to see. The optimal C2 setting
will vary with magnification but is typically 30
–
60%.
e.) If no beam is still visible, try loading a previously saved gun alignment file.
Alignments can be loaded through the “Alignments” control panel, typically
founding under the “Align” tab. Open the flap-out menu, go to the “File” tab,
and select a saved alignment file
. Select “Gun” from the “Available” list by
either double clicking or using the < button, then click apply. If desired,
alignments for other systems/modes of the microscope can also be loaded by
moving the appropriate labels to the “Selected” list and pressing apply.
2.)
One of the control pads (or z axis or tilt buttons) is unresponsive
Re-initialize the pad by unplugging the USB cable from the back of the PC
and re-inserting it. The right plug connects to the right panel, left to the left.
Always note that this occurred in the instrument log
. Reinitializing the
panels may randomly activate functions on that panel, so check for changes
in spot size, diffraction mode, etc. after plugging it back in.