TE70 IR-HEATER
PAGE 24
Calibrate the sensor using the appropriate calibrator on an area of the
surface to be heated, possibly a flat area.
-
Set the calibrator orthogonal to the surface on which the heating
process will be performed.
-
Press and hold down the “PUSH CALIB” button.
-
The calibration process requires from 30 seconds to 1 minute.
-
At the end of the calibration process, the “CALIB” LED stays on
and the control unit will emit 3 beeps to confirm calibration was
performed.
-
Release the “PUSH CALIB” button and put the calibrator back in
its seat.
N.B.: use the calibrator only on surfaces at room temperature; if the
surface to be calibrated had been previously heated, wait until it cools
down before placing the calibrator on the surface. The use of the
calibrator on hot surfaces can damage the calibrator.