TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
76
Procedure for Initial Calibration and Setup
Neumovent Graph Ventilator
Operation
Description
70
Oxigen Valve
Do the same operations than Air valve.
80
Pressure Sensors
The screen will display a list of the calibration values saved in the memory which
correspond to the pressures indicated in the following table:
(zero p) (calibration p.)
(Control Sensor) 0
100 cmH2O
(Graph Sensor)
0
100 cmH2O
(Air Sensor)
0
5 kg/cm2 (71 PSI)
(Oxigen Sensor) 0
5 kg/cm2 (71 PSI
)
the values on the left column, which correspond to zero pressure
Select the desired option with the
“
Selection
” k
ey and press
“
Enter
” to save the
zero
value of the four sensors
a-
Maximum value for the control sensor
(right)
1-
Connect the supply pressures
2- Select the desired option with the
“
Selection
”
key
3-
Connect a
100 cmH2O
pressure at the outlet to the patient.
4-
Save the value with
“
Enter
”.
Presssure Sensors
Derecho
0 0 0
Cm H
2
O
Izquierdo
0 0 0
Cm H
2
O
Aire
0 0 0
Kg./cm
2
Oxigeno
0 0 0
Kg./cm
2
Содержание NEUMOVENT
Страница 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Страница 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Страница 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Страница 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Страница 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Страница 32: ...CHAPTER III TROUBLE SHOOTING ...
Страница 37: ...CHAPTER IV SENSOR VERIFICATION ...
Страница 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Страница 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Страница 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Страница 78: ...CHAPTER VIII CALIBRATION ...
Страница 90: ...CHAPTER IX FINAL CONTROL ...