TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
39
How to Open and Close the
Neumovent Graph
Ventilator
Operation
Description
10
Removal of back cover:
Open the ventilator by removing the eight screws located in the positions indicated
in the figure.
Position of the device for the disconnection of the cables:
- Lay the device down with the face downwards, and place it on foam rubber.
- Lay the back cover down as indicated in the figure, and disconnect the cables.
Adjustment screw and
flange -
A.C. cable
Device
Foam Rubber
Back Cover