TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date: 09/05/06
14
to maintain the adjusted value of positive pressure. It can be programmed with or
without pressure support.
During the inspiratory phase, there is a decrease of the airway pressure proportional
to the demand. During the expiratory phase the airway pressure tend to increase and
becomes higher than the regulated base pressure.
Specific Controls for the CPAP mode:
PEEP/CPAP:
It regulates the airway level of positive pressure.
Combined modes
Group of modes in which the patient has spontaneous ventilation with mandatory inspirations
inserted in synchronized form. Also it comprises modes with spontaneous ventilation and
minimum objectives of tidal volume or minute volume.
The combined modes that may be programmed are:
§
SIMV (VCV) + PSV
Synchronized intermittent ventilation with volume-controlled mandatory inspiration
and spontaneous inspirations with pressure support.
§
SIMV (PCV) + PSV
Synchronized intermittent ventilation with pressure-controlled mandatory inspiration
and spontaneous inspirations with pressure support.
§
MMV + PSV
Mandatory minute ventilation with pressure support. The ventilator has an automatic
control of the pressure support level in order to guarantee minimum minute ventilation
during an eventual decrease of the spontaneous breathing.
§
PSV + VT Assured
Pressure support ventilation with assured tidal volume in case of an eventual
reduction of the breathing effort. In this mode the objective is to guarantee a minimum
tidal volume from a pressure regulated inspiration.
§
Airway Pressure Release Ventilation
(APRV)
It is a mode which ventilates applying periodic switching between two adjustable
levels (P-high and P-lower) of continuous positive airway pressure (CPAP) during
preset periods of time.
Synchronized intermittent ventilation with volume-controlled mandatory
inspiration and spontaneous inspirations with pressure support.
(SIMV [VCV] +
PSV)
Definition and Operative Proceeding:
This mode is a combination of spontaneous
breathing with mechanical ventilation placed synchronically according the patient's
demand. In this synchronized ventilation form, the patient receives during the
mandatory breaths (forced) a preset volume sent with a preset frequency and
inspiratory time. During the spontaneous breathings the patient ventilates with
pressure support.
As in the Volume (VCV) mode, the flow waveform of the mandatory inspiration can be
changed in the course of the ventilation.
Specific Controls for the SIMV [VCV] + PSV mode:
Содержание NEUMOVENT
Страница 1: ...Graph ventilator Technical manual review 04 MAY 2006 ...
Страница 3: ...CHAPTER I GENERAL AND OPERATION FEATURES ...
Страница 12: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 9 ...
Страница 13: ...TECHNICAL AND MAINTENANCE MANUAL Graph VENTILATOR Review 04 Date 09 05 06 10 ...
Страница 30: ...CHAPTER II MAINTENANCE INSTRUCTIONS ...
Страница 32: ...CHAPTER III TROUBLE SHOOTING ...
Страница 37: ...CHAPTER IV SENSOR VERIFICATION ...
Страница 44: ...CHAPTER V EQUIPMENT OPENING AND CLOSURE ...
Страница 60: ...CHAPTER VI DETAIL OF ASSEMBLIES ...
Страница 70: ...CHAPTER VII ELECTRONIC BOARDS DETAIL ...
Страница 78: ...CHAPTER VIII CALIBRATION ...
Страница 90: ...CHAPTER IX FINAL CONTROL ...