Technical data
Proline Promass F 500 EtherNet/IP
268
Shock resistance
Shock, half-sine according to IEC 60068-2-27
6 ms 50 g
Impact resistance
Rough handling shocks according to IEC 60068-2-31
Mechanical load
Never use the transmitter housing as a ladder or climbing aid.
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
Details are provided in the Declaration of Conformity.
16.9 Process
Medium temperature range
Standard version
–50 to +150 °C (–58 to +302 °F)
Order code for "Measuring tube
mat., wetted surface", option
HA,
SA, SB, SC
Extended temperature version
–50 to +240 °C (–58 to +464 °F)
Order code for "Measuring tube
mat., wetted surface", option
SD,
SE, SF, TH
High-temperature version
–50 to +350 °C (–58 to +662 °F)
For nominal diameters DN 25 (1"),
DN 50 (2") and DN 80 (3")
Order code for "Measuring tube
mat., wetted surface", option
TT,
TU
Low-temperature version
–196 to +150 °C (–320 to +302 °F)
NOTICE
Material fatigue due to excessive
temperature differential.
‣
The maximum temperature
differential of the fluid used
must not exceed 300 K.
Order code for "Measuring tube
mat., wetted surface", option
LA