Proline Promass F 500 EtherNet/IP
Diagnostics and troubleshooting
217
12.3 Diagnostic information on local display
12.3.1 Diagnostic message
Faults detected by the self-monitoring system of the measuring device are displayed as a
diagnostic message in alternation with the operational display.
Operational display in alarm condition
Diagnostic message
1
4
3
X X X X
X X X
X
X
X X X X
X X X
X
X
2
S
S
XX
20.50
5
X
i
S801
Menu
Supply voltage
A0029426-EN
1
2
3
4
5
Status signal
Diagnostic behavior
Diagnostic behavior with diagnostic code
Short text
Operating elements
If two or more diagnostic events are pending simultaneously, only the message of the
diagnostic event with the highest priority is shown.
Other diagnostic events that have occurred can be displayed in the
Diagnostics
menu:
• Via parameter
Status signals
The status signals provide information on the state and reliability of the device by
categorizing the cause of the diagnostic information (diagnostic event).
The status signals are categorized according to VDI/VDE 2650 and NAMUR
Recommendation NE 107: F = Failure, C = Function Check, S = Out of Specification, M
= Maintenance Required
Symbol
Meaning
Failure
A device error has occurred. The measured value is no longer valid.
Function check
The device is in service mode (e.g. during a simulation).
Out of specification
The device is operated:
Outside its technical specification limits (e.g. outside the process temperature range)
Maintenance required
Maintenance is required. The measured value remains valid.