CHAPTER 1 INSTALLATION
13
1.2 Installation Conditions
Before installing the instrument, please confirm the following installation
conditions.
1.2.1 Work Power Supply
The requirements for UV2600 work power supply are as follows
:
1
)
Voltage
:
AC100-240V, fluctuation within ±10% is allowed.
2
)
Frequency
:
50Hz or 60Hz
,
fluctuation within ±4% is allowed.
3
)
Power
:
more than 600VA is required. If other devices share the same power
supply,it needs more than 1KVA.
4
)
Grounding wire
:
Grounding resistance must be less than10
Ω
installed as per
relative standards.
1.2.2 Work Table
The requirements for work table are as follows
:
1
)
Work table for UV2600 spectrophotometer should be horizontal and reliable,
withstanding weight of more than 85kg, with a width of more than 800mm, a
depth of more than 750mm.
2
)
If other devices will be put on the work table as well, the area required
should be more.
3
)
Leave space of more
than 200mm at both sides of the instrument. Avoid
instrument touching to the wall directly.
4
)
The height of UV2600 spectrophotometer is about 270mm. For easy
operation, select suitable height of work table.
Содержание UV 2600
Страница 1: ...UV VIS Spectrophotometer UV 2600 Version 1 INSTRUCTION MANUAL UV2600 UV VIS Spectrophotometer ...
Страница 32: ...CHAPTER 4 MAINTENANCE 29 Figure 4 2 move the screw of cuvette holder Figure 4 3 Removal of cuvette holder ...
Страница 42: ...CHAPTER 4 MAINTENANCE 39 Figure 4 12 Install WI lamp Figure 4 13 Shape of WI lamp fuse Bow of fuse Twist ...
Страница 43: ...CHAPTER 4 MAINTENANCE 40 Figure 4 14 Pull out install plug of D2 lamp lead Figure 4 15 Remove install D2 lamp ...