6. Instrument Specifications
92
Instructions for Use for INFINITE M1000 PRO No. 30064852 Rev. No. 1.0
2011-09
6.6 Luminescence
STOP
Caution
Switch on the instrument at least 15 minutes before starting a
luminescence measurement to ensure stable conditions for the
measurement.
Luminescence Detection uses the photon counting technique.
Parameters Characteristics
Wavelength Range
400-700 nm
Dynamic Range
6 decades
Crosstalk (96-well)
≤
0.035 % ATP
Integration Time/well
100 - 20000 ms
Attenuation
No attenuation, Automatic (OD2)
6.6.1
ATP Glow Luminescence
Plate type (number of wells)
384
1536
Total Dispensed Volume[µl]
100
10
Integration Time/well [ms]
1000
1000
ATP Detection Limit [fmol/well]
≤
1
≤
10
6.6.2
Flash Type Luminescence
STOP
Caution
Switch on the instrument at least 15 minutes before starting a
luminescence measurement to ensure stable conditions for the
measurement.
Luminescence Detection uses the photon counting technique.
Parameters Characteristics
Wavelength Range
400 – 700 nm
Measurement Range
> 6 orders of magnitude
Integration Time/well