4. Inserting and Defining Filter Slides
4.3.1
Defining the Filter
Caution
Any changes to the filters in the filter slide are to be carried out by the
service engineer! The instrument is able to recognize predefined filter
slides and you should not attempt to change the filter values.
However, if the filters in the filter slide have been changed (by a service
engineer) or if a new undefined customized filter slide is to be used, the
filter slides need to be defined.
Define a filter as follows:
1. Close
XFLUOR
.
2. Click
the
RdrOLE
icon in the
TECAN
group, and the RdrOLE screen is
displayed.
3. Select
Connect
from the
Instrument
menu. You are now connected to the
GENios.
4. Select
Define Filter
from the
Setup
menu.
5. The following dialog box is displayed showing the filter values for the
excitation and the emission filters:
The Define Filter Dialog Box:
You can choose the options from the right
column of the dialog box for GENIOS
.
6. Enter the new wavelength for each new filter. Ensure that the wavelength
corresponds with the filter position.
7. Accept the new filter values by clicking
OK
. You are now ready to collect
data with the new filters.
8. Exit
RdrOLE
by clicking
Exit
in the
File
menu and restart
XFLUOR
.
9. This completes the installation procedure.
4-6
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
June 2002